Planck LFI CPV: Thermal Susceptibility Test Report
Date Issued
2009
Author(s)
•
•
•
•
Pearson, David
•
•
•
•
GREGORIO, Anna
•
•
•
•
Davis, Richard
•
•
•
Franceschet, Cristian
•
•
Leonardi, Rodrigo
•
Lowe, Stuart
•
•
Meinhold,Peter
•
Mendes, Luis
•
Poutanen, Torsti
•
•
Tomasi, Maurizio
•
•
•
Wilkinson,Althea
•
Zonca, Andrea
Abstract
The SCS PID will be used to characterise the radiometers front-end thermal susceptibility (THF). The temperature of the LFI focal plane (FPU) is changed over 4 values, each step having a duration > 3h, depending on the step. The temperature variations are of about 0.3 K each. The measured receivers output will be characterized as a function of FPU temperature variations.
Rights
open.access
File(s)![Thumbnail Image]()
Loading...
Name
PL-LFI-PST-RP-075_1-1.pdf
Description
Planck LFI CPV: Thermal Susceptibility Test Report
Size
507.11 KB
Format
Adobe PDF
Checksum (MD5)
248845f69a9e972495e3ce4308bae1f0