The X-IFU is the cryogenic spectrometer onboard the future ATHENA X-ray observatory. It is based on a large array of TES microcalorimeters, which work in combination with a Cryogenic AntiCoincidence detector (CryoAC). This is necessary to reduce the particle background level thus enabling part of the mission science goals. Here we present the first joint test of X-IFU TES array and CryoAC Demonstration Models, performed in a FDM setup. We show that it is possible to operate properly both detectors, and we provide a preliminary demonstration of the anti-coincidence capability of the system achieved by the simultaneous detection of cosmic muons.
ATHENA X-IFU Demonstration Model: First Joint Operation of the Main TES Array and its Cryogenic AntiCoincidence Detector (CryoAC)
D'ANDREA, Matteo;ARGAN, ANDREA;BRIENZA, Daniele;LOTTI, Simone;MACCULI, CLAUDIO;MINERVINI, Gabriele;PIRO, LUIGI;Ferrari Barusso, L.;
2022-01-01
Abstract
The X-IFU is the cryogenic spectrometer onboard the future ATHENA X-ray observatory. It is based on a large array of TES microcalorimeters, which work in combination with a Cryogenic AntiCoincidence detector (CryoAC). This is necessary to reduce the particle background level thus enabling part of the mission science goals. Here we present the first joint test of X-IFU TES array and CryoAC Demonstration Models, performed in a FDM setup. We show that it is possible to operate properly both detectors, and we provide a preliminary demonstration of the anti-coincidence capability of the system achieved by the simultaneous detection of cosmic muons.| File | Dimensione | Formato | |
|---|---|---|---|
|
s10909-022-02786-w.pdf
accesso aperto
Licenza:
Non specificato
Dimensione
1.5 MB
Formato
Adobe PDF
|
1.5 MB | Adobe PDF | Visualizza/Apri |
|
Dandrea2022 - JLTD.pdf
accesso aperto
Licenza:
Non specificato
Dimensione
1.13 MB
Formato
Adobe PDF
|
1.13 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.