Please use this identifier to cite or link to this item:
http://hdl.handle.net/20.500.12386/36869
Title: | Performance and uniformity of a kilo-pixel array of Ti/Au transition-edge sensor microcalorimeters | Authors: | Taralli, E. D'ANDREA, Matteo Gottardi, L. Nagayoshi, K. Ridder, M. L. de Wit, M. Vaccaro, D. Akamatsu, H. Bruijn, M. P. Gao, J. R. |
Issue Date: | 2021 | Journal: | REVIEW OF SCIENTIFIC INSTRUMENTS | Number: | 92 | Issue: | 2 | Abstract: | Uniform large transition-edge sensor (TES) arrays are fundamental for the next generation of X-ray space observatories. These arrays are required to achieve an energy resolution $\Delta E$ < 3 eV full-width-half-maximum (FWHM) in the soft X-ray energy range. We are currently developing X-ray microcalorimeter arrays for use in future laboratory and space-based X-ray astrophysics experiments and ground-based spectrometers. In this contribution we report on the development and the characterization of a uniform 32$\times$32 pixel array with 140$\times$30 $\mu$m$^2$ Ti/Au TESs with Au X-ray absorber. We report upon extensive measurements on 60 pixels in order to show the uniformity of our large TES array. The averaged critical temperature is $T_\mathrm{c}$ = 89.5$\pm$0.5 mK and the variation across the array ($\sim$1 cm) is less than 1.5 mK. We found a large region of detector's bias points between 20\% and 40\% of the normal-state resistance where the energy resolution is constantly lower than 3 eV. In particular, results show a summed X-ray spectral resolution $\Delta E_\mathrm{FWHM}$ = 2.50$\pm$0.04 eV at a photon energy of 5.9 keV, measured in a single-pixel mode using a frequency domain multiplexing (FDM) readout system developed at SRON/VTT at bias frequencies ranging from 1 to 5 MHz. Moreover we compare the logarithmic resistance sensitivity with respect to temperature and current ($\alpha$ and $\beta$ respectively) and their correlation with the detector's noise parameter $M$, showing an homogeneous behaviour for all the measured pixels in the array. | URI: | http://hdl.handle.net/20.500.12386/36869 | URL: | https://pubs.aip.org/aip/rsi/article-abstract/92/2/023101/369173/Performance-and-uniformity-of-a-kilo-pixel-array?redirectedFrom=fulltext http://arxiv.org/abs/2102.09348v1 |
ISSN: | 0034-6748 | DOI: | 10.1063/5.0027750 | Fulltext: | open |
Appears in Collections: | 1.01 Articoli in rivista |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
2102.09348v1.pdf | Postprint | 3.09 MB | Adobe PDF | View/Open |
JAP_2021_92.pdf | [Administrators only] | 8.7 MB | Adobe PDF |
Page view(s)
19
checked on Apr 30, 2025
Download(s)
6
checked on Apr 30, 2025
Google ScholarTM
Check
Altmetric
Altmetric