The TES-based Cryogenic AntiCoincidence Detector (CryoAC) of ATHENA X-IFU: A Large Area Silicon Microcalorimeter for Background Particles Detection
Date Issued
2024
Author(s)
•
•
•
•
•
•
Torrioli, G.
•
Chiarello, F.
•
Ferrari Barusso, L.
•
Celasco, E.
•
Gallucci, G.
•
Gatti, F.
•
Grosso, D.
•
Rigano, M.
•
•
Cavazzuti, E.
•
Volpe, A.
Abstract
We are developing the Cryogenic AntiCoincidence detector (CryoAC) of the ATHENA X-IFU spectrometer. It is a TES-based particle detector aimed to reduce the background of the instrument. Here, we present the result obtained with the last CryoAC single-pixel prototype. It is based on a 1 cm2 silicon absorber sensed by a single 2 mm × 1 mm Ir/Au TES, featuring an on-chip heater for calibration and diagnostic purposes. We have illuminated the sample with 55Fe (6 keV line) and 241Am (60 keV line) radioactive sources, thus studying the detector response and the heater calibration accuracy at low energy. Furthermore, we have operated the sample in combination with a past-generation CryoAC prototype. Here, by analyzing the coincident detections between the two detectors, we have been able to characterize the background spectrum of the laboratory environment and disentangle the primary (i.e. cosmic muons) and secondaries (mostly secondary photons and electrons) signatures in the spectral shape.
Volume
214
Issue
3-4
Start page
164
Issn Identifier
0022-2291
Rights
open.access
File(s)![Thumbnail Image]()
Loading...
Name
2024 - J Low Temp Phys - DAndrea et al CryoAC DM127.pdf
Description
Pdf editoriale
Size
1.54 MB
Format
Adobe PDF
Checksum (MD5)