The Demonstration Model of the ATHENA X-IFU Cryogenic AntiCoincidence Detector
Date Issued
2020
Author(s)
•
•
Torrioli, G.
•
•
•
•
•
•
Biasotti, M.
•
Ferrari Barusso, L.
•
Gatti, F.
•
Rigano, M.
•
Volpe, A.
•
Battistelli, E. S.
Abstract
The Cryogenic AntiCoincidence detector (CryoAC) of ATHENA X-IFU is designed to reduce the particle background of the instrument and to enable the mission science goals. It is a 4-pixel silicon microcalorimeter sensed by an Ir/Au TES network. We have developed the CryoAC demonstration model, a prototype aimed to probe the critical technologies of the detector, i.e., the suspended absorber with an active area of 1 cm2; the low energy threshold of 20 keV; and the operation connected to a 50 mK thermal bath with a power dissipation less than 40 nW. Here, we report the test performed on the first CryoAC DM sample (namely, the AC-S10 prototype), showing that it is fully compliant with its requirements.
Volume
199
Issue
1-2
Start page
65
Issn Identifier
0022-2291
Ads BibCode
2020JLTP..199...65D
Rights
open.access
File(s)![Thumbnail Image]()
![Thumbnail Image]()
Loading...
Name
2020 - J Low Temp Phys - DAndrea et al. CryoAC-DM.pdf
Description
[Administrators only]
Size
1.1 MB
Format
Adobe PDF
Checksum (MD5)
d4464c0e24a8b6d5d5014e8c3d791ad7
Loading...
Name
1912.06390.pdf
Description
postprint
Size
1.02 MB
Format
Adobe PDF
Checksum (MD5)
50af1e2884431ca51a8c26dfd752c807