Titolo: | MA_MISS FM Radiometric Calibration | Autori: | DE ANGELIS, Simone FERRARI, MARCO TINIVELLI, PAOLA |
Data pubblicazione: | 2018 | Numero: | EXO-MAM-IAPS-RP-2018-002 | Abstract: | This document reports on the Ma_Miss Flight Model (FM) Radiometric Calibration. The quantities measured by the instrument (i.e. the raw signal converted in digital numbers) are transformed in physical units (the spectral radiance) through the determination of the Instrument Transfer Function (responsivity). The instrument performances have been characterized in terms of (i) acquisitions vs. focal distance, (ii) detector linearity vs. integration time, (iii) the estimate of signal-to-noise ratio, (iv) acquisitions at two different detector temperatures, (v) characterization of optical fiber across TVC and finally (vi) the determination of instrument responsivity. | URI: | http://hdl.handle.net/20.500.12386/33786 | Fulltext: | reserved |